Guidance and manufacturer’s declaration: electromagnetic immunity

The Portable Oxygen Concentrator is intended for use in the electromagnetic environment specified below. The customer or the user of the concentrator should assure that it is used in such an environment.

 

Immunity test

IEC 60601 test

Compliance

Electromagnetic

 

 

level

level

environment/guidance

 

 

 

 

 

 

 

Electrostatic

± 6kV contact

± 15kV contact

Floors should be wood,

 

discharge (ESD)

± 8kV air

± 8kV air

concrete, or ceramic tile.

 

IEC 61000-4-2

 

 

If floors are covered with

 

 

 

 

synthetic material, the

 

 

 

 

relative humidity should

 

 

 

 

be at least 30%.

 

Electric fast

± 2kV for

± 2kV for

Mains power quality

 

transient/burst

power supply

power supply

should be that of a typical

 

IEC 61000-4-4

lines

lines

commercial or hospital

 

 

± 1kV for

± 1kV for

environment.

 

 

input/output

input/output

 

 

 

 

lines

lines

 

 

 

Surge

± 1kV

± 1kV

Mains power quality

 

IEC 61000-4-5

differential mode

differential mode

should be that of a typical

 

 

± 2kV common

± 2kV common

commercial or hospital

 

 

mode

mode

environment.

 

 

 

 

 

 