Guidance and manufacturer’s declaration: electromagnetic immunity
The Portable Oxygen Concentrator is intended for use in the electromagnetic environment specified below. The customer or the user of the concentrator should assure that it is used in such an environment.
| Immunity test | IEC 60601 test | Compliance | Electromagnetic | |
|
| level | level | environment/guidance | |
|
|
|
|
|
|
| Electrostatic | ± 6kV contact | ± 15kV contact | Floors should be wood, | |
| discharge (ESD) | ± 8kV air | ± 8kV air | concrete, or ceramic tile. | |
| IEC |
|
| If floors are covered with | |
|
|
|
| synthetic material, the | |
|
|
|
| relative humidity should | |
|
|
|
| be at least 30%. | |
| Electric fast | ± 2kV for | ± 2kV for | Mains power quality | |
| transient/burst | power supply | power supply | should be that of a typical | |
| IEC | lines | lines | commercial or hospital | |
|
| ± 1kV for | ± 1kV for | environment. | |
|
| input/output | input/output |
|
|
|
| lines | lines |
|
|
| Surge | ± 1kV | ± 1kV | Mains power quality | |
| IEC | differential mode | differential mode | should be that of a typical | |
|
| ± 2kV common | ± 2kV common | commercial or hospital | |
|
| mode | mode | environment. | |
|
|
|
|
|
|