TAG

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

37.50

 

 

 

 

 

 

 

 

 

 

 

 

ppm CH4

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.00

 

 

 

 

Range: 1

50.00

 

 

 

 

 

 

 

 

 

 

Failures:

 

 

 

 

No

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Maintenance-Requests: No

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Temperature:

 

20.0 C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.0

 

 

 

 

 

100.0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Operation:

 

 

 

Ready

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Display

 

Status...

 

Main...

 

Channel

 

Calib...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

F2

 

F3

F4

 

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TAG

 

 

 

 

 

 

 

 

 

 

37.50 ppm

 

 

 

 

 

 

 

 

 

 

 

-- Analyzer Module Calibration --

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Calibration procedure status...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Start zero calibration procedure !

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Start span calibration procedure !

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Time & Date:

 

 

 

12:50:55 June 16, 1998

 

 

 

 

 

 

 

 

Valve position:

 

 

 

 

 

 

Samplegas

 

 

 

 

 

 

 

 

Expected zero gas:

 

 

 

 

 

 

0.00 ppm

 

 

 

 

 

 

 

 

Expected span gas:

 

 

 

 

 

 

50.00 ppm

 

 

 

 

 

 

 

 

Span gas name:

 

 

 

 

 

 

 

Methane

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measure

 

Status...

 

Channel

 

Back...

 

Valves...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

 

F2

 

F3

 

F4

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TAG

 

 

 

 

 

 

 

 

 

 

37.50 ppm

 

 

 

 

 

 

 

 

 

 

 

 

-- Set Gas Valves --

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Flow zero gas !

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Flow span gas !

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Flow sample gas !

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Flow test gas !

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Close all valves !

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Valve position:

 

 

 

 

 

 

Samplegas

 

 

 

 

 

 

 

 

Operation status:

 

 

 

 

 

 

 

Ready

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measure

 

Status...

 

Channel

 

Back...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

 

 

F2

 

 

F3

 

 

F4

 

 

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4.10Flow Test Gas

Change to the menu "Analyzer Module Calibration"

F5

Press

Starting this function will switch an external valve and flow the MLT with test gas.

Change to the menu "Set Gas Valves"

F5

Press

Change to the line "Flow test gas !"

Press or as often as necessary to get the line "Flow test gas !" white on black.

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98

NGA 2000

4 - 59