ATEX Instruction Manual

ETC01035

January 2004 MLT 2 / BINOS® 100 F

Emerson Process Management Manufacturing GmbH & Co. OHG 1 - 5
Depending on which system is used, diffe-rent parameters apply:
Application Mode of Operation
Single Enclosure
(volume approx. 56 l)
Dual Enclosure (volume
approx. 112 l)
Zone 1 Prepurge Pha se
for measuring Duration: 10 min 14 min
flamm a ble ga ses Purge gas consumpt i on: 1,75 Nm³/h 1,92 Nm³/h
maximum overpressure: 95 mbar 95 mbar
Measuring Mode
Internal overpressure: 90mbar 90 mbar
Flow rate (leakage rate): <= 3 l / m i n <= 4,5 l/ min
Zone 1 Prepurge Pha se
for measuring Duration: 4 m i n 6 min
non-flammable ga ses Purge gas consumpt i on: 6,8 Nm³/h 9,2 Nm³/h
maximum overpressure: 25 mbar 25 mbar
Measuring Mode
Internal overpressure: 14 mbar 14 mbar
Flow rate (leakage rate): <= 3 l / m i n <= 4,5 l/ min
Application Mode of Operation Single Enclosure
(volume approx. 56 l)
Dual Enclosure (volume
approx. 112 l)
Zone 2 Prepurge Pha se
for mea suri n g Duration: 5 m in 17 min
non-flammable ga ses Purge gas consumpt i on: 8 Nm³/h 7 Nm³/h
maximum overpressure: 25 m bar 25 mbar
Measuring Mode
Internal overpressure: 1 mbar 1 mbar
Flow rate (leakage rate): <= 22Nm³/h <= 22 Nm³/h
Minimum overpressure in all enclosures:1 mbar against external atmosphere and50 Pa against containment system when measuring flammable gases