Model NGA2000 NDIR

Instruction Manual

748332-F April 2003

CAUTION

HAND INJURY HAZARD

Dropping the front panel of the Platform while hand or fingers are inside either case handle can cause serious injury.

CAUTION

PARTS INTEGRITY

Tampering with or unauthorized substitution of components may adversely affect safety of this product. Use only factory approved components for repair.

CAUTION

OVERBALANCE HAZARD

This Analyzer Module may tip instrument over if it is pulled out too far and the Platform is not properly supported.

CAUTION

POSSIBLE CELL DAMAGE

Apply leak test liquid to cell or detectors only as a last resort.

NOTICE

Software compatibility is necessary for all NGA2000 components in your system to work together. The version of your Platform's software must be equal to or greater that the version of any other module(s) for successful compatibility. If it is not, contact Rosemount Analytical at 800-433-6076 to order software upgrade kit PN 657150 for the Platform.

You can locate the version of each NGA2000 component as follows:

Platform Controller Board

Turn power ON. The display will show "Control Module V2. ...". This is the software version.

Analyzer Module

Located on the right side of the Analyzer Module case.

I/O Module

Located on the backplane connector of the module. If no label is present, the module is Version 2.0.

Rosemount Analytical Inc. A Division of Emerson Process Management

Preface P-3