Addendum
5.Connect the instrument sample inlet tub- ing to the Vaisala Pressure Inlet port.
6.Enable the Barometric Pressure Com- pensation and the Use Local Reading op- tions via the Expert Controls – Analyzer Module Diagnostics – Barometric Pres- sure Compensation menus of the instru- ment. The instrument will automatically detect and report the ‘Presence’ of the Pressure Compensation Option.
Circuit adjustment can be made using Barometric Pressure menu display, or an external Digital voltmeter.
Display Method
With no sample inlet lines connected and the instrument exhaust ported to atmosphere, adjust R1 on the 661380 Interface board until the Barometric Pressure reading on the
Model NGA2000 PMD, NDIR
Barometric Pressure menu reads the correct ambient air pressure.
DMM Method
Connect the DMM (+) to TP1 on the 661380 Interface board,
V.The instrument is now calibrated at 1 ATM.
For applications where the instrument is to be used above 5400 Feet (MSL) an alternate transducer option is available from the factory. This version consists of a Vaisala model PTB100B transducer