The "See Manual" indicator displays two fault types:
Hardware Fault. The "See Manual" indicator flashes off and on. Briefly pressing the Temperature Setting button briefly produces a pattern on the three Temperature Setting Indicators, representing a fault code. Refer to § 5.2 for fault code definitions.
Microprocessor Fault. The "See Manual" indicator is on continuously.
3.4.43 The
If a heated circuit is connected and the
Refer to § 5 for trouble shooting.
3.4.5AMBIENT TEMPERATURE SENSORMeasures ambient air temperature for controlling the heating of the heated wire breathing circuits. Refer to § 3.5.1 for more information.
3.4.6CHAMBER DETECTION SENSORA thermistor is embedded in the third pin in the