4MAINTENANCE & CLEANING PROCEDURES

4.1CLEANING

The following is the recommended cleaning procedure for the MR810 Respiratory Humidifier. Solutions listed below have been tested to ensure that no damage will occur with the metal and plastic components of the humidifier. Cleaning should be performed as required.

1.

Disconnect the humidifier from the electrical outlet.

 

2.

Clean the humidifier with one of the following using a damp cloth:

 

 

Normal dishwashing detergent

 

 

Isopropyl alcohol

 

 

4

3.

Wipe the humidifier clear of any cleaning residues before use.

 

 

 

 

 

Caution

DO NOT immerse the humidifier in any liquid. Using other solutions may damage the humidifier.

4.2ANNUAL MAINTENANCE

To ensure that the humidifier is safe and effective in operation the following maintenance shall be carried out ANNUALLY.

4.2.1VISUAL CHECK

Check the humidifier for physical damage as follows:

1.Check the power cable for damage, replace if necessary (§ 6.6).

2.Check the heater-wire adaptor for kinks, abrasions or damage to the connector. Replace if necessary (§ 6.6).

3.Check the heater plate for deep scratches, replace if necessary (§ 6.5.3).

4.Check the humidifier case for cracks and ensure case screws are fitted and correctly tightened (the three screws under the humidifier). Check the finger guard is fitted and prevents contact with the heater plate during normal operation.

4.2.2ELECTRICAL SAFETY CHECK

Check that the humidifier is electrically safe to use. Refer to § 7.

4.2.3PERFORMANCE TEST PROCEDURE

Carry out the performance test procedure described in § 8.

MR810 Technical Manual Revision C REF 185042601