5.2SEE MANUAL CODES
When the See Manual Indicator is illuminated the humidifier will switch off the heater plate and heater- wire. Pressing the Temperature Setting button displays the error code on the Temperature Setting Indicators, as explained in Table 3.
Table 3: Indicator codes and their meaning
Temperature | See |
| Fault Description |
|
| Section |
| |
Setting | manual |
|
|
|
|
|
|
|
Indicators | Indicator |
|
|
|
|
|
|
|
|
| Microprocessor failure, replace PCB. |
| 6.4 |
|
| ||
|
|
|
|
|
|
| ||
|
| Button fault. Check buttons are correctly seated. |
| 6.7.1 |
|
| ||
|
|
|
|
|
| |||
|
| Heater plate primary thermistor fault (short or open | 6.5.4 |
|
| |||
|
| circuit). Check the thermistor is working correctly, |
|
|
| |||
|
| replace if necessary. |
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
| Heater plate | secondary thermistor | fault (short | or | 6.5.4 |
|
|
| 5 | |||||||
|
| open circuit). Check the thermistor is working |
|
| ||||
|
|
|
|
| ||||
|
| correctly, replace if necessary. |
|
|
|
|
| |
|
|
|
|
|
| |||
|
| Ambient thermistor fault (short or open circuit), | 6.6 |
|
| |||
|
| check |
|
|
|
| ||
|
|
|
|
|
| |||
|
| Chamber Sense thermistor fault (short or open | 6.6 |
|
| |||
|
| circuit), check |
|
|
| |||
|
| necessary. |
|
|
|
|
|
|
|
|
|
|
|
| |||
|
| 6.4 |
|
| ||||
|
| faulty, service or replace PCB as necessary. |
|
|
|
| ||
|
|
|
|
|
| |||
|
| Heater control triac error | 6.4 |
|
| |||
|
| circuit or heater plate triac open/short circuit). |
|
|
| |||
|
| Service triac circuitry, replace PCB if necessary. |
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
| Clip Circuit error. | Service | the | 6.4 |
|
| |
|
| hardware clip circuit, replace PCB if necessary. |
|
|
|
|
Failed factory production testing. Device should be returned to your Fisher & Paykel Healthcare representative.
Temperature Setting Indicator description
Indicator off
Indicator flashing
Indicator on
MR810 Technical Manual Revision C REF 185042601 | 17 |