6.5.7INSTALLING THE HEATER PLATE

Heater plate spring

 

Figure 6.9: Heater plate assembly

 

1.

Ensure the heater plate springs are located in their respective location holes on the heater plate.

6

2.

Pull the heater plate harness through the hole in the case top (Figure 13).

 

3.Carefully orient the case upper over the heater plate springs, then screw the three mounting screws back into place.

6.6REPLACING THE POWER CABLE OR HEATER-WIRE ADAPTOR

Note: The spare heater-wire adaptor, part number 043 042 322, includes the cable clamp and collet for the mains cable and heater-wire adaptor.

Table 10: Part numbers and colour codes of Fisher & PaykelHealthcare power cables

Model

Supply

Power Cable

Phase

Neutral

Earth

 

Plug

 

Voltage

Part Number

Colour

Colour

Colour

 

 

MR810ALU

230 V~

095 428 323

Brown

Blue

Green / Yellow

 

Schuko

MR810AGU

 

 

 

 

 

 

 

MR810AFU

 

 

 

 

 

 

 

MR810ARU

 

 

 

 

 

 

 

MR810ANU

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MR810AEA

230 V~

095 428 317

Brown

Blue

Green / Yellow

 

Australasian

 

 

 

 

 

 

 

 

MR810AEK

230 V~

095 428 569

Brown

Blue

Green / Yellow

 

UK

 

 

 

 

 

 

 

 

MR810AEU

230 V~

095 428 856

Brown

Blue

Green / Yellow

 

No Plug

 

 

 

 

 

 

 

 

MR810ADU

230 V~

095 428 854

Brown

Blue

Green / Yellow

 

Danish

 

 

 

 

 

 

 

 

MR810JHU

115 V~

095 428 322

Brown or

Blue or

Green / Yellow

 

USA

MR810JSU

 

 

Black

White

 

 

Right angle

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MR810GJU

100 V~

095 428 869

Brown or

Blue or

Green/Yellow

 

Japanese

 

 

 

Black

White

 

 

 

 

 

 

 

 

 

 

 

MR810 Technical Manual Revision C REF 185042601

27