D.3 ELECTROMAGNETIC IMMUNITY
Guidance and manufacturer’s declaration – electromagnetic immunity
The MR810 is intended for use in the electromagnetic environment specified below. The customer or the user of the MR810 should assure that it is used in such an environment.
Immunity Test | IEC60601 | Compliance Level | Electromagnetic environment - |
| test level |
| guidance |
Electrostatic | ± 6 kV contact | ± 6 kV contact | Floors should be wood, concrete or |
discharge (ESD) | ± 8 kV air | ± 8 kV air | ceramic tile. If floors are covered |
IEC |
|
| with synthetic material, the relative |
|
|
| humidity should be at least 30 %. |
Electrical fast | ± 2 kV for power | ± 2 kV for power | Mains power quality should be that |
transient/burst | supply lines | supply lines | of a typical commercial or hospital |
IEC | ± 1 kV for | ± 1 kV for | environment. |
| input/output lines | input/output lines |
|
Surge | ± 1 kV differential | ± 1 kV differential | Mains power quality should be that |
IEC | mode | mode | of a typical commercial or hospital |
| ± 2 kV common | ± 2 kV common | environment. |
| mode | mode |
|
Voltage dips, short | < 5 % UT | < 5 % UT | Mains power quality should be that |
interruptions and | (> 95 % dip in UT for | (> 95 % dip in UT for | of a typical commercial or hospital |
voltage variations on | 0,5 cycle) | 0,5 cycle) | environment. If the user of the |
power supply input | 40 % UT | 40 % UT | MR810 requires continued |
lines | operation during power mains | ||
IEC | (60 % dip in UT for 5 | (60 % dip in UT for 5 | operation, it is recommended that |
| cycles) | cycles) | the MR810 shall be powered from |
| 70 % UT | 70 % UT | an uninterruptible power supply or |
| battery. | ||
| (30 % dip in UT for | (30 % dip in UT for |
|
| 25 cycles) | 25 cycles) |
|
| < 5 % UT | < 5 % UT |
|
| (> 95 % dip in UT for | (> 95 % dip in UT for |
|
| 5 seconds) | 5 seconds) |
|
Power Frequency | 3 A/m | 3 A/m | Power frequency magnetic fields |
(50/60 Hz) magnetic |
|
| should be at levels characteristic of |
field |
|
| a typical location in a typical |
IEC |
|
| commercial or hospital |
|
|
| environment. |
Note: UT is the a.c. mains voltage prior to application of the test level.
D
MR810 Technical Manual Revision C REF 185042601 | 41 |