2SPECIFICATIONS
2.1 | MECHANICAL |
|
|
| 2 | ||
| Dimensions: | 94 x 154 x 135 mm (without chamber fitted). | |
| Weight: | 1.7 kg without chamber. |
|
|
| 2.0 kg with MR290 filled with water. |
|
2.2 | ELECTRICAL |
|
|
| Supply Voltage: | 230 V~ |
|
|
| 115 V~ |
|
|
| 100 V~ |
|
| Supply Frequency: | 50 or 60 Hz |
|
| Supply Current: | 0.8 A maximum at 230 V~ |
|
|
| 1.6 A maximum at 115 V~ |
|
|
| 1.8 A maximum at 100 V~ |
|
| 22 V~ at 30 W |
| |
| Heater Plate: | 150 W |
|
| Heater Plate Thermal Cutout: | Operates at 93 ± 5 °C |
|
2.3 | PERFORMANCE |
|
|
Note: Performance results obtained at 23 °C ambient temperature.
2.3.1HEATER-WIRE MODE PERFORMANCE
Table 1: Heater-wire mode performance for each temperature setting
Temperature | Constant flow range |
| Delivered patient temperature (°C) |
|
Setting | (L/min) |
|
| |
|
|
| ||
Low | 5 to 60 |
| 26 to 29 | |
|
|
|
| |
Medium | 5 to 60 |
| 30 to 33 | |
|
|
|
| |
High | 5 to 30 |
| 33 to 36 | |
|
|
|
|
|
Results obtained using an RT308 oxygen therapy breathing circuit.
2.3.2NON HEATER-WIRE MODE PERFORMANCE
Table 2: Non heater-wire mode performance for each Temperature Setting
Temperature | Constant flow range | Delivered patient |
|
Setting | (L/min) | temperature (°C) |
|
Low | 5 to 60 | 23 to 25 |
|
|
|
|
|
Medium | 5 to 40 | 25 to 27 |
|
|
|
|
|
High | 5 to 30 | 28 to 32 |
|
|
|
|
|
Results obtained using an RT307 oxygen therapy breathing circuit. |
| ||
MR810 Technical Manual Revision C REF 185042601 | 7 |