Choice will depend upon application.
Please contact your local Fisher & Paykel Healthcare representative for recommendations.
TEMPERATURE DISPLAY
Displays the saturated gas temperature (the lower of the airway and chamber temperatures in °C) delivered to the patient. This display will normally show the chamber temperature (around 37 ± 0.5 °C for invasive mode, and 31 ± 0.5 °C for noninvasive mode).
By pushing and holding the mute button for one second, the chamber outlet temperature then the airway temperature is displayed. The display will then revert to normal operation.
LOW TEMPERATURE ALARM
An audible alarm and flashing temperature display showing 35.5 °C or lower (in invasive mode only). Caused by cold/drafty conditions or very high or low gas flows.
If the circumstances causing the low humidity alarm cannot be changed then the audible alarm acts as a reminder that the patient is receiving inadequate humidity and may require further intervention to maintain airway clearance.
HIGH TEMPERATURE ALARM
A flashing temperature display showing 41 °C or higher. The humidifier will discontinue heating of the chamber and circuit until the temperature decreases to within normal limits.
MODE BUTTON
This button switches between invasive and noninvasive mode.
Invasive ModeThis mode is for patients with bypassed airways. The humidifier delivers gas as close to body temperature saturated (37 °C, 44 mg/L) as possible.
Under cold or drafty conditions the chamber temperature may drop as low as
35.5°C in order to maintain a dry breathing circuit.
WARNING: Ensure that invasive mode is set for patients that have bypassed airways
Noninvasive ModeThis mode is for patients receiving face mask or headbox therapy, and delivers a comfortable level of humidity.
ON/OFF BUTTON
The humidifier will power ON if this button is held down briefly, but must be held down for one second to turn the humidifier off. The humidifier will always default to invasive mode when it is turned on.
ROUTINE MAINTENANCE
Refer maintenance to qualified service personnel. A full technical description including routine maintenance and service data is contained in the Technical Manual which is available from your supplier or Fisher & Paykel Healthcare (REF 185041340)