
Millenium™ RO System by GE Osmonics
3.1Daily
WARNING:
Verification of the absence of all sanitizers, cleaners, disinfectants and chlorine/chloramines must be performed before RO product water is used for any patient related uses.
1.Connect the RO unit to the required utilities (feed water, drain, and power).
2.Ensure that all the valves in the water supply line to the RO are open.
3.Ensure that all
4.Ensure that the product rinse line is connected to the RO, with the other end in a drain.
5.Move the inlet water valve switch to the open position. When water flows to drain, the RO pump power switch can be moved to the on position.
6.The "Percent Rejection" monitor should indicate 95% rejection within 2 minutes of
7.The TDS monitor should stabilize within 2 minutes of
WARNING:
When the RO is used for dialysis service, a test for disinfectant residual MUST be performed.
8.Record data on a daily log sheet.
WARNING:
Any data that does not meet minimum standards set at your facility must be reported to the medical director. The RO should not be placed in service until readings are acceptable.
9.The Millenium™ RO is now ready for connection to the dialysis machine.
3.2Shut Down Procedure
1.Turn off the RO pump power switch.
2.Turn off the inlet water valve switch.
3.Turn off the inlet water supply.
14693e - 31 Mar 03 | 3 - 3 | System Operation |