Millenium™ RO System by GE Osmonics
4.4Millenium™ Disinfection Indications
DETERMINATION OF MICROBIAL CULTURING AND DISINFECTION FREQUENCY IS THE RESPONSIBILITY OF THE OPERATING FACILITY. The following is recommended for critical applications where bacteria and endotoxins are a major concern, such as for Hemodialysis.
Disinfection should occur when bacterial growth is above the limits set by the facility, or when over 200 CFU/ml (maximum AAMI standard) is detected. Routine disinfection at lower CFU counts will inhibit colonization of bacteria in the system and the distribution plumbing.
Rapid
CAUTION:
Glutaraldehyde and Actril™ are not recommended disinfectants for the Millenium™.
4.4.1Formaldehyde Disinfection Preparation
CAUTION:
New replacement thin film membranes must be rinsed to drain with normal pressures for a minimum of six (6) hours before first exposure to formaldehyde, or irreversible damage may occur. No mandatory
WARNING:
Always wear rubber gloves, apron and facemask. Use proper ventilation during disinfection and
Materials Required:
•800 ml. Formaldehyde, 37%
•2 gallons pre treated or RO water
•clean wide mouth container with volume of 2 ½ gallon or more
•formaldehyde test kit
•product rinse line (clear hose with male QD fitting on one end, open on other)
•“DO NOT USE” warning labels or signs (not provided by Osmonics)
•cleaning / disinfect log sheets
•most current log sheets for water quality comparison
•proper protective body wear
WARNING:
14693e - 31 Mar 03 | 4- 18 | Routine Maintenance |