4.6.2BIOSAN™ Storage Procedure

WARNING:

It is recommended that gloves (latex/vinyl and eye protection be worn when handling BIOSAN™.

1.Disinfect and rinse the RO with an acceptable disinfectant.

2.Follow the BIOSAN™ procedure in this chapter.

3.Do not rinse the RO out after BIOSAN™ has been circulated in the RO.

4.Disconnect the disinfect line, waste line, feed water line, power and product line.

5.Coil the hoses and power cord, and cover the RO with a large plastic bag to protect the RO from dust.

CAUTION:

The Millenium™ RO may be stored for up to two weeks in BIOSAN™. After two weeks, if longer storage is needed, the Millenium™ RO may be rinsed and repacked with BIOSAN™.

WARNING:

After storage, the RO unit must be completely rinsed of BIOSAN™, and a bacterial culture performed and interpreted before use. Perform AAMI analysis if it has been one year or more since last done.

14693e - 31 Mar 03

4- 29

Routine Maintenance