4.6.2BIOSAN™ Storage Procedure
WARNING:
It is recommended that gloves (latex/vinyl and eye protection be worn when handling BIOSAN™.
1.Disinfect and rinse the RO with an acceptable disinfectant.
2.Follow the BIOSAN™ procedure in this chapter.
3.Do not rinse the RO out after BIOSAN™ has been circulated in the RO.
4.Disconnect the disinfect line, waste line, feed water line, power and product line.
5.Coil the hoses and power cord, and cover the RO with a large plastic bag to protect the RO from dust.
CAUTION:
The Millenium™ RO may be stored for up to two weeks in BIOSAN™. After two weeks, if longer storage is needed, the Millenium™ RO may be rinsed and repacked with BIOSAN™.
WARNING:
After storage, the RO unit must be completely rinsed of BIOSAN™, and a bacterial culture performed and interpreted before use. Perform AAMI analysis if it has been one year or more since last done.
14693e - 31 Mar 03 | 4- 29 | Routine Maintenance |