Descriptions of Refrigerator Parts and
Their Functions
1)Control Panel
The control panel on the Haier blood bank shows the actual storage temperature inside the refrigerator. You can use the push buttons on the panel to adjust the temperature set point and test the alarm functions.
2)Light Power Switch
You can use the switch to turn on and off the interior lights of the refrigerator.
3)Condensation Heater Control
You can use this switch (located inside the back of the refrigerator behind a cover) to turn on the heater that reduces the condensation build up. “1” means power on and “0” means power off.
4)Battery Power Switch
The switch controls the rechargeable battery that powers the alarm system in case of the loss of main power. “1” is on and “0” is off.
5)Door Lock
You can use this feature to lock the refrigerator.
6)Door Handle
For ease of opening and closing the door.
7)Casters
The casters underneath the unit provide the convenience of moving the unit around.
8)Automatic Temperature Recorder (optional)
The refrigerator is equipped with an automatic temperature recorder. Its paper chart runs for a full turn in 7 days, so the storage temperature can be permanently recorded. The chart paper must be changed once every 7 days.
9)Leveling Leg
The unit must be level to operate. You can adjust the screws on the leveling leg underneath the refrigerator to adjust the height.
10)Shelves
The drawers and shelves are designed to conveniently store and unload the blood samples.
11)Interior Lights
15 Watts light hulb.
12)Door Ajar Alarm
When the door is open for 10 minutes, the door ajar alarm will sound. You should reduce the open door time to stabilize the refrigerator temperature.
Maintenance
Warning
Before any inspection or maintenance work is performed, the refrigerator's power plug should be disconnected from the power supply socket. This is to prevent any potential electrical shock or injury. During the maintenance work, do not breathe the dust and aerosols near the unit; they might be harmful to your health.
Cleaning the refrigerator
1.The refrigerator should be cleaned once a month. Regular cleaning can keep the unit looking new. 2.Use a dry, soft cloth to clean the dust off the interior and exterior. If necessary, use a soft cloth with a solution of water and mild detergent to wipe the unit.
3.After the cleaning, use a dry cloth to wipe off any solution residue off the surfaces.
4.Do not pour water directly into the refrigerator. By doing so, the water can damage the insulation materials and cause problems.
5.Parts in the refrigeration system for this refrigerator are completely sealed. They do not require any lubricants.
6.The air filter for the condenser should be cleaned at least once within a 90 day period. The procedure is as follows.
a.Remove the front louver panel for the refrigerator.
b.Remove the air filter located in front of the condenser; wash it in water. c.After cleaning the filter, install it in front of the condenser and the louver panel.
Changing the light bulb starter
The florescent light bulb is located in the upper front area of the refrigerator. When changing the light bulb, please perform the following steps.
1.Unplug the refrigerator from the power supply.
2.Remove the left panel and the cover for the electrical control box. 3.Remove the defective starter.
4.Install a new starter. Make sure it is securely in position. 5.Install the cover for the electrical box and the left panel. 6.Plug the power plug back into the system.
Purchasing chart paper for temperature chart recorder
The chart recorder papers are imported chart papers designed for the recorder. The papers that are supplied with the unit usually last about half of a year. When you are running short of the chart paper, please contact Haier to purchase more. The lead time to deliver the charts is 15 days after receiving the payment. Please call Haier for details.
Purchasing backup battery for temperature recorder
You can purchase a high quality battery from a store for the same size and voltage (9V). You may also call Haier to purchase a high quality replacement battery. The lead time to deliver the battery is 15 days after receiving the payment. Please call Haier for details.
Changing fuse breaker
If there is a burnt out fuse on a PC board, please contact Haier's after sales support department.