2.3.9
2.3.9.2Image Tab
The Image tab is for setting operating condition with respect to image observation.
Fig.
(1)ABCC Window (SE/BSE)
If the results of ABCC are not adequate, you can change the reference brightness and contrast. Adjustable range is
(2)FAST SCAN AVERAGING Window
Select number of frames to be averaged for TV1, 2 and Fast1, 2 speeds. Recommended numbers are 4 for TV1 and 8 or 16 for TV2, 2 for Fast1 and 4 for Fast2. The higher the number of frames, the better the attainable image quality. However, the higher number results in longer persistence time. Select an optimal value for the present observed image.
NOTICE: The number of frames to be averaged is set for each screen mode (Full, Small or Dual) independently.
2 - 55