For Safe Handling

Precautions for Proper Operation

Precautions for the Proper Operation of the S-3400N

About Electricity

(a)Single-phase AC 100 V(±10%), 50/60 Hz, 2 kVA (for main unit).When the supplied power line voltage is 100 V, the instrument works by direct connection.

When the supplied power line voltage is other than 100 V (200 V or others), use a step down transformer to step the power line voltage down to 100 V. Refer to Chapter 1 (Specifications and Installation Requirement) for the step down transformer requirement.

(b)Connect the instrument to ground based on the standard of your country to prevent electric shock hazards due to the high voltage. Otherwise, you could suffer an electric shock or a malfunction could occur in the instrument.

Avoid sharing the ground terminal with other power equipments. Be sure to ground the instrument independently. The image trouble and the malfunction of the equipment might be caused when sharing.

Handling Volatile Solvents

When handling a volatile solvent in cleaning components of this system, take precautions on the following points:

(a)Volatile solvent should be handled at a place which is well ventilated and is well away from flames.

(b)For handling volatile solvents, ventilation equipment and protective gear should be provided at the customer’s discretion. The inhaling of a volatile solvent in excessive amounts can results in respiratory difficulties. If volatile solvent is swallowed or it gets onto the skin or into the eyes, symptoms of polyneuritis including anesthesia and ataxia (difficulty in walking) may develop.

SAFETY - 12