For Safe Handling

Precautions on Proper Operation (cont’d)

Transportation and Relocation of Instrument

(a)The transport of the system, which carries considerable risk, should never be attempted by a customer. Any system transport should be performed with the assistance and supervision of Hitachi-approved engineers.

(b)Do not lift the instrument by holding the table. The strength of table fitting is not sufficient for bearing the weight of display unit, approximately 120 kg. Should the table be lifted, the display unit might slip off and crash. Hence, it is recommended to remove the table and transport the display unit independently when moving the instrument.

Ambient Temperature and Humidity

Even when the system is not running, the room in which the system is installed should be maintained under the following environmental conditions:

Room temperature: 15to 30

Humidity: 70or lower (no condensation)

If the system is not used for a long time, the specimen chamber should be evacuated instead of being exposed to the atmosphere.

SAFETY - 20