2.2Testing Communications

Background

Once you connect power and the SCT or SFC to the transmitter, you are

 

 

 

 

 

 

 

ready to test communications with the transmitter.

Procedure

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The procedure in Table 2 outlines the steps using an SFC for initiating

 

 

 

 

 

 

 

communications with an ST 3000 transmitter without an assigned tag

 

 

 

 

 

 

 

number.

 

 

 

 

 

 

 

Table 2

 

Testing Communications with Transmitter.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Step

 

Press Key

 

 

 

 

Read Display or Action

Description

 

 

 

 

 

 

 

 

 

 

 

 

 

1

 

 

 

 

 

 

Slide power switch on left side of SFC to

SFC runs its self check and displays

 

 

 

 

 

 

 

 

ON position.

 

 

 

 

initial prompt.

 

2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

If this prompt appears, transmitter is

 

 

 

 

 

 

 

 

P

U

T

L

O O P

I N

M A N

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

in Analog mode of operation. This is

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

the factory default mode of

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

operation setting. Put your control

 

 

 

 

 

 

 

 

 

 

 

 

 

OR

 

 

loop in the manual mode of

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

operation before initiating SFC

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

communications. Note that you must

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

do this separately through the

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

receiving device in the loop.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

If this prompt appears, transmitter is

 

 

 

 

 

 

 

 

 

D E

– X M T R

P R E S S I D

 

 

 

 

 

 

 

 

 

 

 

in Digital (DE) mode of operation.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3

 

 

DE READ

 

 

 

 

 

 

 

 

 

 

Be sure any switches that may trip

 

 

 

 

 

T

A

G

N O .

 

 

 

 

 

 

 

A

 

 

 

T

R

I

P S

S E

C U R

E D ? ?

 

alarms or interlocks associated with

 

 

 

 

ID

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

analog loop are secured or turned

 

 

 

 

 

 

 

 

 

 

 

OR

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

off. Go to Step 4.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Go to Step 5

 

 

This prompt does not appear for

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

transmitters operating in DE mode.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

See DE transmitter display response

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

in Step 5.

 

4

 

 

NON-VOL

 

 

Confirms that “TRIPS” are secured. Go

Required for transmitters operating

 

 

 

 

ENTER

 

 

 

to Step 5 for display response.

in analog mode only.

 

 

 

 

(Yes)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Continued on next page

2/05

ST 3000 Release 300 Installation Guide

7