Specifications
Technical | |||
Detector type | Germanium | InGaAs | Filtered InGaAs |
Optical interface/connectors | Universal 2.5 mm/1.25 mm1 | Universal 2.5 mm/1.25 mm1 | Universal 2.5 mm/1.25 mm1 |
Wavelength range | 780 to 1600 nm | 780 to 1650 nm | 780 to 1650 nm |
Wavelength settings | 780 to 1600 nm, in 1 nm steps | 780 to 1650 nm, in 1 nm steps | 780 to 1650 nm, in 1 nm steps |
Programmable wavelengths | 5 presets (customized) | 5 presets (customized) | 5 presets (customized) |
Calibrated wavelengths | 850, 980, 1300, 1310, | 850, 980, 1300, 1310, | 850, 980, 1300, 1310, |
| 1490, 1550 nm | 1490, 1550, 1625 nm | 1490, 1550, 1625 nm |
Power range | |||
Display range | |||
Max. input power | +13 dBm | +16 dBm | +27 dBm |
Measurement units | dB/dBm/W | dB/dBm/W | dB/dBm/W |
Absolute uncertainty2 | ± 0.2 dB (±5%) | ± 0.2 dB (±5%) | ± 0.2 dB (±5%) |
Linearity3 | ± 0.06 dB | ± 0.06 dB | ± 0.06 dB |
Tone detection | 270 Hz, 1 kHz, 2 kHz | 270 Hz, 1 kHz, 2 kHz | 270 Hz, 1 kHz, 2 kHz |
Auto λ mode4 | Yes | Yes | Yes |
Multi λ mode4 | TwinTest/serial TripleTest | TwinTest/serial TripleTest | TwinTest/serial TripleTest |
Notes:
1.Optional 1.25 mm UPP accessory adapter is available
2.Under these reference conditions: 20 dBm (CW), 1310 nm ±1 nm, 23°C ±3K, 5 to 75% relative humidity 9 to 50 μm test fiber with ceramic DIN connector
3.
4.With JDSU light sources
General
Storage andinterface |
|
Data storage | 100 results |
Data download capability | micro USB interface for PC transfer (option) |
Power supply |
|
Dry batteries | 2x Mignon (AA) 1.5 V |
Rechargeable batteries | 2x Mignon (AA) NiMH 1.2 V |
AC operation (optional) | via micro USB and |
Battery life | ≥ 200 hr |
| Automatic power off after 20 minutes |
Environmental conditions |
|
EMI/ESD | CE compliant |
Recommended calibration interval | 3 years |
Operating temperature | |
Storage temperature | |
Dimensions (H x W x D) | 30 x 80 x 150 mm (1.2 x 3.1 x 5.9 in) |
Weight | 200 g (0.45 lb) |