NI USB-6008/6009 User Guide and Specifications 20 ni.com
Figure 11. Connecting a Reference Single-Ended Voltage Signal
Digital Trigger
When an AI task is defined, you can configure PFI 0 as a digital trigger
input. When the digital trigger is enabled, the AI task waits for a rising or
falling edge on PFI 0 before starting the acquisition. To use ai/Start Trigger
with a digital source, specify PFI 0 as the source and select rising or falling
edge.
Analog OutputThe NI USB-6008/6009 has two independent AO channels that can
generate outputs from 0–5 V. All updates of AO lines are software-timed.
Analog Output Circuitry
Figure 12 illustrates the analog output circuitry for the NI USB-6008/6009.
Figure 12. Analog Output Circuitry
AI
NI USB-6008/6009
GND
Voltage
Source
REF(+) REF(–)
12-Bit
DAC
Output
Buffer
50 Ω
AO
+5 V
GND