8.BASIC PROGRAMMING

8.4 Chart Speed Programming

Program the chart speed before operation.

Press SHIFT and 1 simultaneously in the

CHART

operation screen to display the [Chart speed] programming screen.

For the remote contacts (option), read Section13.3.

0000mm/h can be set for RD2800. When 0000mm/h is set, chart is fed at the speed of 12.5mm/h.

Programming range

0001 to 0600 (mm/h)

or

0001 to 0200 (mm/h)

1 Programming Mode



Speed unit (Fixed) Cursor Chart speed (4 digits programming) H/M

Character display

Programming lamp

Speed unit

(H or M)

CursorMemory colon

2

Programming Flow Chart

 

 

 

 

 

<Example> From 20 mm/h to 50 mm/h

 

 

 

 

 

 

 

 

Operation screen

 

DISP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SHIFT

1

 

 

[Check]

 

 

 

CHART

 

 

 

 

 

 

 

 

 

 





 

 

 

 

 

 

Speed unit

 

Chart speed

 

 

 

 

 

 

 

 

 

 

SET

 

 

 

 

[To be ready for programming]

 

END

 

 

 

 

 

 

 

 

 

 





 

 

 

 

 

 

 

 

Cursor

 

 

 

 

 

 

 

 

 

 

 

 

 

[Programming chart speed]

 

0

0

5

0

 

CLEAR

CLEAR

DATA I

CLEAR

 

 

 

 

 

 

 





 

 

 

 

 

 

 

 

Cursor

 

 

 

 

 

 

 

 

 

 

ENTRY

 

 

 

 

YES

 

Error message appears?

 

 

 

Keys other than

 

 

 

NO

 

 

 

 

 

 

SHIFT

 

 

 

 

 

 

 

 

 

Stored

 

 

 

 

Changing the speed unit (h/m)

Move the cursor to the character display (a digit on the left) then press

/or % to switch “M” and “H” alternatively.

Programmed chart speed

Cursor and programming lamp light.

Cursor moves rightward when pressing a numeric key.

Cursor moves rightward.

Cursor moves leftward.

Program the chart speed within 0001 to 0600 mm/h or 0001 to 0200 mm/min.

An error occurs if a value exceeds the programming range or any space or decimal point is entered.

The new programmed value is stored into memory.

Memory colon blinks and the programming change mark is printed.

Remarks 1

If 150 mm/h or more is

programmed for the chart speed

 

 

 

No printing, except time line, data printing and programming change mark, is executed.

Remarks 2 Influence to periodic data printing

When changing the chart speed, the programmed parameters for [periodic data printing] (Section 11.7) is cleared. (Default: Not programmed)

-44-

Page 48
Image 48
Omega RD200, RD2800 manual Chart Speed Programming,    , , 