General

 

Range/Sensor type

Pt100 to IEC 751 (ITS 90 refers). –199.99 to 849.99°C.Ro=

 

100Ω. Two input channels, each 3 or 4 wire connection with

 

automatic recognition (with manual overrride).

Overall Instrument

±0.02°C ± 1 digit for range -200°C to +500°C.

Accuracy (4 wire)

±0.005% reading ± 1 digit for range 500°C to 850°C.

Overall System

Better than ±0.04°C with TM-l250 precision probe accuracy

 

(4 wire) from -50°C to +250°C with 5 point NAMAS

 

calibration.

Linearisation

 

Conformity

±0.01°C.

Stability

 

(vs ambient

 

temperature)

0.0025°C/°C ambient change.

Warm-up Time

Negligible under normal ambient conditions. Allow 5 to 10

 

minutes for full stability unless stored at low temperature,

 

then allow 30 minutes minimum.

Pt100 Sensor Current

0.5mA nominal.

Display Resolution

0.01°C, K, °F, Ω.

Measurement Units

°C, °F, K, Ω (user selectable).

Measurement Modes

Input A or B or A-B (differential). Null facility in A-B mode.

Custom Calibration

 

(via PC software)

Up to 10 calibration values can be allocated to (via PC

 

software) channels A and B. Values are retained in non-

 

volatile memory until replaced by user.

Null Function

Corrects differential temperature readout between the two

 

Sensors to zero.

Sensor Lead

 

Resistance

25Ω each lead maximum.

 

Page 17

Supply

Internal rechargeable batteries. Mains 110/240V 50/60Hz,

 

adapter included. Battery charge life up to 12 hours with full

 

charge dependent on pattern of usage.