Pentair IC20, IC40 Troubleshooting Continued, IntelliChlor Installation and User’s Guide, Problem

Models: IC20, IC40

1 32
Download 32 pages 27.03 Kb
Page 29
Image 29

23

Table 1:Troubleshooting (Continued)

Problem

Possible Cause

Corrective Action

 

 

 

Red ADD salt light is

Pool water needs salt. No chlorine is being

Add salt as described on pages 13, 14, and

on.

produced.

15.

 

 

 

Yellow LOW salt light is

Not enough salt in pool.

Add salt to pool to achieve 3200 ppm to 3500

on.

Heavy Rainfall.

ppm. See page 13, 14, and 15.

 

Leak in pool.

 

 

 

 

Green GOOD salt light

Too much salt in pool. May cause corrosion on

Dilute pool water by draining some water, then

is flashing

other pool equipment.

filling with fresh water

 

 

 

PWR light is red

Call for service.

Unit requires attention.

 

 

 

Cell light is red

Water is below 52°F.

The water temperature must be above 52°F to

 

 

produce chlorine.

 

 

 

Cell light does not

Chlorine Production set to 00%.

Adjust CHLORINE PRODUCTION to desired

come on.

 

percentage.

 

 

 

 

Insufficient water flow. Cell is plugged with

Remove obstruction and/or clean cell. See

 

debris, pump has lost prime.

page 10. Prime pump if necessary.

 

 

 

 

Salt level below 2500 ppm.

Add salt as described on pages 13, 14,

 

 

and 15.

 

 

 

CELL light is flashing

Cell has calcium build-up and requires

Refer to Maintenance Procedure for acid

green.

cleaning.

wash/cleaning. See "User Maintenance,"

 

 

on page 9.

 

 

 

LIFE light is flashing

Cell life has achieved 10,000 hours, life is

Prolong cell life by keeping salt at GREEN

green.

limited.

level, and minimize BOOST modes.

 

 

 

Flow light is red

Pump fails to provide sufficient water flow.

Check for correct operation of the pump, i.e.,

 

 

loss of pump prime or clogged strainer

 

 

baskets.

 

 

 

 

Closed valves.

Check and correct all valve alignments.

 

 

 

 

Dirty filter.

Follow filter cleaning procedures.

 

 

 

 

Obstruction in the Chlorinator cell.

Remove cell for inspection. Follow cleaning

 

 

procedures. See "Electrolytic Cell Cleaning,"

 

 

on page 10.

 

 

 

IntelliChlor unit does

Fuse in Power Center is open.

Replace AC fuse, located at bottom of Power

not have green power

 

Center.

light.

 

 

 

 

 

 

No AC power to Power Center.

Verify time clock is providing 110 VAC or 220

 

 

VAC to Power Center when active.

 

 

 

 

Transformer leads not wired correctly in Power

Verify transformer leads wired to AC source by

 

Center.

referring to wiring diagram decal on inside of

 

 

Power Center cover.

 

 

 

IntelliChlor Installation and User’s Guide

Page 29
Image 29
Pentair IC20, IC40 Troubleshooting Continued, IntelliChlor Installation and User’s Guide, Problem, Possible Cause