3-6

6. Change

the

filament

voltage

supply,

if necessary,

depending

on

the

type of

focus.

 

 

 

 

 

A

fine

focus

tube should have a filament

voltage

of

8

volts,

and broad

and

normal

focus

tubes

require

10

volts.

On

later

tubes the

type of

focus

is

indicated,

and in

other

cases consult

the

tube instruction

card.

To

change

the filament voltage, switch off

the

unit,

open

the

right-hand

side

panel

by

loosening

the

two

Allen

screws.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The

panel

is

hinged at the bottom. Behind

the

front

panel

is

a

switch

marked BF

(broad

focus)

and

 

NF FF (normal/fine focus). Select as required.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Close

the

cover.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

7. Close the air inlet valve, and turn on the water supply.

8. Turn the kV and mA controls to the minimum position, and operate the GENERATOR I push-button.

9. When an X-ray

tube

is

new, or

has

not

been used for a

week or

more, the following procedure

should be

followed:

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Check

that the

mA

meter reading is

stable, then

slowly increase the kV control

to

30

kV,

over

a period

of one

minute.

Slowly

increase

the

mA

control

watching

for signs

of instability

of

the

mA

reading.

DO NOT EXCEED THE RATED kV AND mA VALUES, as shown in Appendix 1; disregard the instruction card values (X-ray diffraction tubes).

Slowly increase the kV to

the

required operating high-voltage,

at

the

same

time

adjusting

the

mA reading

so that the maximum power

to

the tube does not exceed the

rated

limits.

If the

mA

meter

reading shows

any sign of instability the kV

should be reduced, and held at

a

lower value

for

about

a

minute before

attempting to advance the

kV

control again.

 

 

 

 

 

 

 

 

3.7.POSITIONING THE TUBE SHIELD

 

The tube shield can be placed in one of

three

orientations on

top

of the generator, i.e. so that

the

X-ray

 

 

 

beams make angles of either 90° or 45°

or

to the

front

of

the

generator.

 

 

 

 

 

 

 

 

 

 

 

 

However, the rotation of the tube shield

must

not

 

be

 

more

than

90°

in

either

direction from its

original

 

 

 

position. Since the line and point focus beams

are

 

90°

apart

it

 

is

possible to place the tube shield with

either

 

beam

in any position

depending

on

the

use

of

the

generator,

 

or

personal

preference.

 

 

 

 

 

 

 

 

 

The tube shield is fixed to the table

top

by means of four Allen screws. Remove the screws and

 

rotate

the

 

tube

 

shield carefully, simultaneously ensuring that the cables and

water

hoses

inside

the

cabinet

are

 

not

unduly

 

tensioned. For this purpose open the left-hand

side

panel

of

the

generator cabinet, and

observe

the

hoses

and

 

cable.

Refix

the

tube

shield on the table

top

with

the

screws,

and

close

the

side

panel.

 

 

 

 

 

 

 

 

3.8.

DIFRACTOMETRY

 

OPERATION

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The generator will normally be used to obtain

the

simultaneous (photographic)

registration

of

reflections

from

 

a

specimen,

therefore,

stabilisation

of

the

X-ray tube voltage

is

not

necessary.

 

 

 

 

 

 

 

 

 

 

 

A

PE

1612

a.c.

voltage

stabiliser

can

be

used

in

 

the

 

mains

supply

to

the

generator,

when

an

improved

stabilisation

 

is required for more accurate powder diffractometry.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

This voltage

stabiliser

is

specially

designed

to provide a stable mains voltage to small units, such

as

the

PW

1720

 

X-ray generator. Variations in the mains

voltage

 

from

-15% to + 10% will be stabilised so that the output

 

 

voltage variation is better than 0,1%. A

full

specification

is

given

in

the

operation

manual of

the

PE 1612

a.c.

 

voltage

stabiliser.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

A PW 1050 vertical goniometer with stepping motor or with synchronous motor can be installed on top of the generator, and two mounting holes are provided in the top cover, see figure 3.4. Refer to the PW 1349 “Diffractometer kit” instruction manual, for alignment information.