OPERATING INSTRUCTIONS
STEP 3
Connecting the remote control cable
4
5
6
7
2.6
Insert the protective cover into the lock housing during use of the device.
Check the control cable and source assembly connectors using the NO GO gauge according to the daily inspection section.
Remove the protective rubber cap from the remote control connector.
Slide the remote control connector assembly's collar back and open the jaws to expose the male portion of the control cable connector
(i.e. the
Press back the
the source assembly connector with a
Release the locking pin of the source assembly connector and check that the connection is secure.
Use the NO GO gauge to check the gap between the joined connectors during the daily inspection.