TECHNICAL SPECIFICATIONS
Source output
|
| At 1m per Ci (37GBq) |
| At 1ft per Ci (37GBq) |
| ||||
0.125R/hr | 1.25mSv/hr |
| 1.3R/hr/Ci | 13.0mSv/hr | |||||
0.203R/hr | 2.03mSv/hr |
| 2.2R/hr/Ci | 22.0mSv/hr | |||||
| 0.48R/hr | 4.80mSv/hr |
| 5.2R/hr/Ci | 52.0mSv/hr | ||||
| 1.30R/hr | 13.0mSv/hr |
| 14.0R/hr/Ci | 140mSv/hr | ||||
Selected attenuation data |
|
|
|
|
|
|
| ||
Material | Approximate |
| Approximate half value thickness |
| |||||
| material |
|
|
| inches(mm) |
|
|
| |
| density | ||||||||
Concrete | 2.35g/cm3 | 1.140 (29.0) | 1.180 | (30.0) | 1.700 | (43.2) | 2.400 | (61.0) | |
Aluminum | 2.65g/cm3 | – |
| 1.100 | (27.0) | – |
| – |
|
Steel | 7.80g/cm3 | 0.170 (4.3) |
| 0.315 | (8.0) | 0.512 | (13.0) | 0.827 | (21.0) |
Lead | 11.34g/cm3 | 0.047 (1.2) |
| 0.039 | (1.0) | 0.200 | (5.1) | 0.500 | (12.7) |
Tungsten | 17.80g/cm3 | – |
| 0.032 | (0.8) | 0.130 | (3.3) | 0.310 | (7.9) |
DU | 18.70g/cm3 | – |
| – | 0.050 | (1.3) | 0.270 | (6.8) | |
Operating distance
The distance between the remote control and the exposure head is determined by summing the length of the remote control conduits plus the total length of source guide tubes used, and normally should not exceed 47ft (14.2m).
Standard remote controls:
25ft (7.6m), 35ft (10.7m), 50ft (15m).
Standard source guide tubes (projection sheaths):
7ft (2.1m) set of three source guide tubes: 21ft total (6.3m).
NOTE
The total length of source guide tubes utilized must be shorter in length than the remote controls to ensure the source assembly will project into the working position.
e.g. When using 25ft (7.6m) remote controls, a maximum of three
7ft (2.1m) source guide tubes can be used with a combined length of 21ft (6.4m).
Conversely, when using 35ft (10.7m) remote controls, a maximum of four 7ft (2.1m) source guide tubes can be utilized.
When using 50ft (15m) remote controls, a maximum of six 7ft (2.1m) source guide tubes can be utilized. This limitation is due to the weight and friction of remote control cable length.
1.3