Sigma LBA-712, LBA-710, LBA-714PC, LBA-300, LBA-708, LBA-500PC manual Ieee 488.1 Command Support, Unl

Models: LBA-500PC LBA-708 LBA-710 LBA-700 LBA-712 LBA-714PC LBA-400 LBA-300

1 281
Download 281 pages 14.04 Kb
Page 211
Image 211

Appendix A Remote Command/Error Message Operation

The LBA-PC can be controlled as a remote device via GPIB. For the most part, communications between the LBA-PC and the host controller will follow the data format and coding protocols outlined in the IEEE 488.1 and 488.2 standards.

A.1 IEEE 488.1 Command Support

The LBA-PC in combination with an appropriate National Instruments GPIB board conforms to the IEEE

488.1 standard according to the following Subset Codes: SH1, AH1, T6, L4, SR1, RL2, PP0, DC1, DT0, E2, C0

These Subset Codes are defined in the IEEE 488.1 standard and describe interface capabilities of the LBA-PC.

The following multiple line interface message commands are supported by the LBA-PC:

GTL

Go To Local (0x01). When the LBA-PC detects the GTL message it returns to

 

local command mode.

LLO

Local Lockout (0x11). When the LBA-PC detects the LLO message it enters what

 

is called the lockout state. This is equivalent to entering the operator password

 

in the Options Password menu item. While in the lockout state access is denied

 

to many of the LBA-PC menu items. When the LBA-PC exits the lockout state

 

access to all menu items is restored.

MLA

My Listen Address (0x20 to 0x3e).

MSA

My Secondary Address (0x60 to 0x7e).

MTA

My Talk Address (0x40 to 0x5e).

PPC

Parallel Poll Configure (0x05). Handled by GPIB board and NI-488.2M.

PPD

Parallel Poll Disable (0x70 to 0x7e). Handled by GPIB board and NI-488.2M.

PPE

Parallel Poll Disable (0x60 to 0x6e). Handled by GPIB board and NI-488.2M.

PPU

Parallel Poll Unconfigure (0x15). Handled by GPIB board and NI-488.2M.

SPD

Serial Poll Disable (0x19).

SPE

Serial Poll Enable (0x18).

UNL

Unlisten (0x3f).

UNT

Untalk (0x5f).

The following IEEE 488.1 commands are not supported:

DCL, GET, SDC, TCT