Sigma LBA-300, LBA-710, LBA-714PC, LBA-712, LBA-708, LBA-500PC, LBA-700 Key Type Value Description

Models: LBA-500PC LBA-708 LBA-710 LBA-700 LBA-712 LBA-714PC LBA-400 LBA-300

1 281
Download 281 pages 14.04 Kb
Page 258
Image 258

Key

Type

Value Description

 

 

 

FrameNumber

I

Frame number

 

 

-1 = gain frame

 

 

0 = reference frame

 

 

1 to n. ‘n’ is the size of the frame buffer.

 

 

default = current frame

 

 

 

Row

I

1 to h. ‘h’ is the frame height.

 

 

default = current cursor location

 

 

 

Returns:

RCR FrameNumber=f;Row=r;#dn..n(DAW)( DAW)…(DAW)(DAW)

Where:

f

=

frame number

r

=

which row is being returned

#= pound symbol

d

=

number of digits to follow in n..n

n..n

=

number of data words (i.e. row length)

(DAW)

=

data word is two 8-bit data bytes, low byte followed by high byte

Each data word is a two’s complement fixed point value in one of the following formats:

siiiiiii ifffffff

LBA-300PC

-256 to 255.9921875

siiiiiii iiifffff

LBA-400PC

-1024 to 1023.96875

siiiiiii iiiiifff

LBA-500PC

-4096 to 4095.875

siiiiiii ifffffff

LBA-708PC

-256 to 255.9921875

siiiiiii iiifffff

LBA-710PC

-1024 to 1023.96875

siiiiiii iiiiifff

LBA-712PC

-4096 to 4095.875

siiiiiii iiiiiiif

LBA-714PC

-16384 to 16383.5

Where:

 

 

 

s

=

sign bit

 

I

=

integer

 

f

=

fraction

 

Operator’s Manual