
Basic Installation
TLA 700 Series Installation Manual 3
Table 1 lists the operating requirements for the logic analyzers. Refer to the
TLA 700 Series Logic Analyzer User Manual for a complete specification list.
Table 1: Mainframe operating requirements
Requirement TLA 704 Color Portable mainframe TLA 711 Color Benchtop mainframe
Source voltage 90–250 VRMS, 45–66 Hz, continuous range
100–132 VRMS, 360–440 Hz, continuous range 90–250 VRMS, 45–66 Hz, continuous range
100–132 VRMS, 360–440 Hz, continuous range
Steady state input current 6 ARMS, maximum 15 ARMS, maximum at 90 VAC
6.3 ARMS, maximum at 207 VAC
Maximum power consumption 600 W 1350 W line power
Temperature
(No media in floppy disk drive) +5° C to +50° C (+41° to +122° F), 15° C/hr
(55° F/hr) maximum gradient, noncondensing
(derated 1° C (34° F) per 1000 ft (304.8 m)
above 5000 ft (1524 m))
+5° C to +50° C (+41° to +122° F), 15° C/hr
(55° F/hr) maximum gradient, noncondensing
(derated 1° C (34° F) per 1000 ft (304.8 m)
above 5000 ft (1524 m))
Humidity
(No media in floppy disk drive) 20% to 80% relative humidity, noncondensing.
Maximum wet bulb temperature: +29° C
(84° F) (derates relative humidity to ~22% @
50° C (+122° F))
20% to 80% relative humidity, noncondensing.
Maximum wet bulb temperature: +29° C
(84° F) (derates relative humidity to ~22% @
50° C (+122° F))
Altitude To 10,000 ft (3,040 m)
(derated 1° C (34° F) per 1000 ft (304.8 m)
above 5000 ft (1524 m))
To 10,000 ft (3,040 m)
(derated 1° C (34° F) per 1000 ft (304.8 m)
above 5000 ft (1524 m))
Operating Requirements