EDIT Menu

14.Select Execute from the sub-menu.

If the number of points in the waveform between the vertical bar cursors in the copy source is the same as the value set for Interval, the copied waveform will be displayed repeating at the interval designated with Interval. In the example shown in Figure 3 -64, both the waveform point size and the Interval value are set to 500.

0

500

999

Figure 3 64: Waveform Copying (number of points in waveform = Interval value)

If the number of points in the waveform between the vertical bar cursors in the copy source is less than the value set for Interval, data at level 0 (in other words, a straight line) will be added at the end of the waveform. Figure 3 -65 shows an example in which the waveform point size has been set to 500 and the Interval has been set to 600. Data at level 0 has been added from point 500 to point 599; the next waveform begins from point 600.

AWG2021 User Manual