EDIT Menu

DC Even Component

Odd Component

Figure 3 "151: Fundamental, Odd, and Even Components

Shifting Magnitudes. Here is the procedure for using the Shift Mag item to shift the magnitude to the limit level.

1.Select Limiter from the bottom menu.

2.Press the CURSOR button on the front panel to select the left/right vertical bar cursor. Using the numeric keys or the general purpose knob, set the domain to be used when shifting the magnitude.

3.Press the VALUE button on the front panel. Using the general purpose knob or the numeric keys, set the limit level for the magnitude. The limit level is displayed at the top the screen.

4.Select Shift Mag from the side menu. The entire area within the left and right vertical bar cursors will shift so the signal with the maximum magni- tude between the cursors becomes the limit level.

AWG2021 User Manual

3 "209