MODE Menu
Setting Trigger Parameters for an External Trigger (Gate)
The external trigger (gate) signal is input from the TRIGGER INPUT connector on the front panel. The maximum input voltage is ±10 V into a 1 MW input impedance, and the maximum input voltage is 5 VRMS into a 50 W input impedance.
Use the Slope or Polarity, and Level items in the side menu of the MODE menu to set trigger (gate) parameters for an external trigger (gate) signal.
Slope. This menu item sets the slope for the external trigger signal. Press the Slope button in the side menu and select either Positive or Negative. When Positive is selected, the output is triggered at the rising edge of the external trigger signal; when Negative is selected, the output is triggered at the falling edge of the external trigger signal.
Polarity (Gated mode). This menu item sets the polarity for the gate that outputs the waveform or sequence with an external gate signal. Press the Polarity button in the side menu and select either Positive or Negative. When Positive is selected, the waveform or sequence is output while the level of the gate signal is higher than the gate level set with the side menu Level item. When Negative is selected, the waveform or sequence is output while the level of the gate signal is lower than the set gate level.
Level. This item sets the trigger (gate) level for an external trigger (gate) signal. Press the Level button in the side menu, then set the level with the numeric keys or the general purpose knob. The trigger (gate) level can be set in the range ±5.0 V to 5.0 V in steps of 0.1 V.
Figure 3
PositiveGoing Edge | NegativeGoing Edge |
Trigger
Level
Can be
Adjusted
Vertically
Trigger Slope Can be Positive or Negative
Figure 3 $168: Slope and Level Controls
AWG2021 User Manual | 3 | $241 |