UTILITY Menu

Figure 3 ￿186: Menu Displayed When RS232C is Selected

Baudrate

Sets the transmission rate. The transmission rate of

 

300, 600, 1200, 2400, 4800, 9600, and 19200 can be

 

selected with the general purpose knob. Set this

 

parameter to the same value as set on the computer.

Parity

Sets the error detection method. The general purpose

 

knob is used to select parity of None, Odd, or

 

Even. Set this parameter to match the connected

 

computer's parity.

Data Bits

Selects 7 or 8 data bits. Set this parameter to match

 

the connected computer's data bits.

Stop Bits

Selects 1 or 2 stop bits. Set this parameter to match

 

the connected computer's stop bits.

Flagging

Selects None, Soft, or Hard. This item is used to

 

announce that the input buffer is full. Until the

 

transmission allowed instruction is issued, the other

 

device stops the data transfer.

AWG2021 User Manual