Appendix A: Options and Accessories

Clock

Data

Raise Time

Fall Time

 

 

 

 

 

 

 

 

 

 

Rise Time / Fall Time : measure the time for the waveform to go from 10% to 90% of its peak value ; must be than 1ns.

Level High : -0.735V ￿ -1.02V

Level Low: -1.6V ￿ -1.95V

Skew : Measure at 50% level of the waveform; must be than 1ns.

Figure A￿5: Output Waveform

If a cable is used, these waveforms have transmission distortion. It is necessary to latch the data with a clock before using the waveform in actual circuits at the cable receiving side (user side) and to reproduce the waveform. Delay the clock with the delay line in order to reproduce the data reliably (see Figure A-6).

Data 12

 

12

 

12

 

 

 

Buffer

Latch

 

 

 

 

 

 

 

 

 

 

 

 

 

Delay Line

Clock

Buffer

4 ns

Figure A￿6: Data Latching

A￿6