Appendix B: Performance Characteristics

Warranted Characteristics

This section will describe the warranted characteristics of the AWG2021. These can be divided into two main categories: electrical characteristics and environmental characteristics.

Performance Conditions The electrical characteristics are valid under the following conditions:

1.The instrument must have been calibrated at an ambient temperature between +20_ C to +30_ C.

2.The instrument must be in an environment whose limits are described in Environmental Characteristics.

3.All tolerance limits apply after a 20 minute warm up and an execution of the self calibration.

4.The instrument is operating at an ambient temperature between +10_ C to +40_ C, unless otherwise noted.

Items marked with ** are tested in the Performance Verification (Appendix C)

Table B"3: Electrical Characteristics

Characteristics

Description

 

 

 

 

Clock Generator

 

 

 

 

 

** Accuracy

 

 

** +15_ C to +30_ C

0.005%

 

 

 

 

 

+10_ C to +40_ C

0.01%

 

 

 

Skew between CH1 and CH2

When Option 02 (second channel) installed within 4 ns

 

 

 

 

Filter

 

 

 

 

Cutoff frequency

-3 dB point

 

 

 

 

 

1 MHz

within 20%

 

 

 

 

 

5 MHz

within 20%

 

 

 

 

 

20 MHz

within 20%

 

 

 

 

 

50 MHz

within 20%

 

 

 

Main Output

 

 

 

 

** Amplitude

 

 

DC Accuracy

 

 

 

0.05 V ￿ 0.5 V

±(0.5% of Amplitude + 5 mV)

 

 

 

No Offset, at 1 MHz clock

 

 

 

 

 

 

** 0.501 V ￿ 5 V

±(1% of Amplitude + 25 mV)

 

 

 

No Offset, at 1 MHz clock

 

 

 

 

B"6