Appendix E: Functional Operation Summary
AW2021 | AW2021 |
TRIR | TRIR |
INPUT | INPUT |
OUT
Trigger Source
Figure E5: Connection Example 2
When there is only one trigger source output, connect to both AWG2021 instruments using the same minimum possible length 50 W cable and a T adapter.
For Triggered or Burst mode, output synchronized with the trigger is obtained by applying the external trigger at the same time to the AWG2021 instruments.
When continuous synchronized output is necessary, use Gated mode. SYNC output is obtained from when the gate signal goes True until it goes False. After the gate signal goes false, to output again, it is necessary to reset the output with the side menu STOP item.
Output synchronized with the clock. Connect the clock output of the master AWG2021 to the clock input of the slave AWG2021 with a 50 W cable and set the slave's clock source to External.
Figure
E6 | AWG2021 User Manual |