12
3.1.2.3. Connecting the Status Outputs
If you wish to utilize the analyzer’s status outputs to interface with a device that accepts logic-level digital inputs,
such as programmable logic controller (PLC) chips, you can access them through a 12 pin connector on the
analyzer’s rear panel labeled STATUS.
EMITTER BUS
FOR PINS 1-8
STATUS
1 2 3 4 5 6 7 8 D +
SYSTEM OK
HIGH RANGE
CONC VALID
ZERO CAL
SPAN CAL
DIAGNOSTIC
MODE
LOW SPAN
Figure 3-5: Status Output Connector
NOTE
Most PLC’s have internal provisions for limiting the current the input will draw. When
connecting to a unit that does not have this feature, external resistors must be used to
limit the current through the individual transistor outputs to 50mA (120 Ω for 5V
supply).
Table 3-3: Status Output Signals
PIN # STATUS CONDITION (ON = CONDUCTING)
1 SYSTEM OK ON if no faults are present.
2 CONC VALID ON if concentration measurement (NO, NO2 or NOx) is valid.
OFF any time the hold-off feature is active.
3 HIGH RANGE ON if unit is in high range of the Auto Range Mode.
4 ZERO CAL ON whenever the instrument is in ZERO point calibration mode.
5 SPAN CAL ON whenever the instrument is in SPAN point calibration mode.
6 DIAG MODE ON whenever the instrument is in diagnostic mode.
7 LOW SPAN CAL ON when in low span calibr ation (optional equipment necessary)
8 Unused
D EMITTER BUS The emitters of the transistors on pins 1-8 are tied together.
Unused
+ DC POWER + 5 VDC, 300 mA (combined rating with Control Output, if used).
Digital Ground The ground level from the analyzer’s internal DC power supplies