Teledyne API - Model 200EH/EM Operation Manual Getting Started
29
STEP 2 – activate O2 sensor stability function
To change the stability test function from NOx concentration to the O2 sensor output, press:
SETUP X.X STABIL_GAS:O2
NO NO2 NOX O2 ENTR EXIT
Press EXIT 3
times to return
to SAMPLE
menu
SAMPLE A1:NXCNC 1=100PPM NO X=XXX.X
< TST TST > CAL SETUP
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG ALRM EXIT
SETUP X.X ENTER PASSWORD:818
8 1 8 ENTREXIT
SETUP X.X 0) DAS_H OLD_OFF=15.0 M inutes
<PREV NEXT> JUMP EDIT PRNT EXIT
SETUP X.X 2) STABIL_GAS=NOX
<PREV NEXT> JUMP EDIT PRNT EXIT
SETUP X.X STABIL_ GAS:NOX
NO NO2 NOX O2 ENTREXIT
Continue pressing NEXT until ...
NOTE
Use the same procedure to reset the STB test function to NOx when the O2 calibration
procedure is complete.
04521C (DCN5731)