68 Model 7500 Instruction Manual
Teledyne Analytical Instruments
8. TROUBLESHOOTING8.1 Error message
If errors occur, the following contents are displayed.
Error display Error contents Probable causes
Error No. 1 Sector motor rotation detector signal is faulty. • Infrared ray light source is faulty.
• Sector motor rotation is faulty or stopped.
• Motor rotation detector circuit is faulty.
• Amplier circuit is faulty.
Error No. 3 A/D conversion signal is faulty. • Circuit is faulty.
Error No. 4 Zero calibration is not within the allowable
range.
• Zero gas is not supplied.
• Zero point is deected much due to a dirty cell.
• Detector is faulty.
Error No. 5 A amount of zero calibration (indication value) is
over 50% of full scale.
Error No. 6 Span calibration is not within the allowable
range.
• Span gas is not supplied.
• Calibrated concentration setting does not
match cylinder concentration.
• Zero calibration is not performed properly.
• Span is deected much due to dirty cell.
• Detector sensitivity is deteriorated.
Error No. 7 A amount of span calibration (difference
between indication value and calibrated
concentration value) is over 50% of full scale.
Error No. 8 Measured values uctuate to much during zero
and span calibration.
• Calibration gas is not supplied.
• Time for supplying calibration gas is not short.
Error No. 9 Calibration is abnormal during auto calibration. • Error corresponding to No. 4 to No. 8 occurred
during auto calibration.
Error No. 10 Output cable connection is improper. • Wiring is detached between analyzer and
interface module.
• Wiring is disconnected between analyzer and
interface module.