Teledyne API – T101 Operation Manual Operating Instructions
131
4.8.2.2. Editing DAS Data Channels
DAS configuration is most conveniently done through the APICOM remote control
program. The following list of key strokes shows how to edit using the front panel.
Main Data Acquisition Menu
Edit Dat a Channe l Menu
SAMPLE RANGE = 500.0 PPB H2S =XXX.X
< TST TST > CAL SETU P
SETUP X.X 0)CONC: ATIMER, 2, 4032, R
PREV NE XT INS DEL EDIT PRNT EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SAMPLE ENTER SETUP PASS : 81
8
8 1 8 ENTR EXIT
EXIT will return to the
previous SAMPLE
display.
SETUP X.X DATA ACQUISITIO N
VIEW
EDIT EXIT
Moves the
display up &
down the li st of
Data Channels
Inserts a new Data
Channe l into the list
BEFORE the Channel
curre ntly bei ng displa yed Deletes The Data
Channel currently
being di splaye d
Exports the
confi guratio n of all
data cha nnels to
RS-232 i nterfac e.
Exits to the Main
Data Acquisition
Menu
SETUP X.X NAME:CONC
<SET SET> EDIT PRNT EXIT
Moves the display
between the
propert ies fo r th is d ata
channel.
Reports the conf iguratio n of curr ent
data cha nnels to the RS-23 2 ports.
EXIT returns to the
previo us Menu
To edit th e chann el name, s ee next sequ ence.
When editing the data channels, the top line of the display indicates some of the
configuration parameters. For example, the display line:
0) CONC: ATIMER, 4, 800
translates to the following configuration:
Channel No.: 0
NAME: CONC
TRIGGER EVENT: ATIMER
07266B DCN6485