Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485)
A-35
58 Sample pressure “Hg
60 Internal box temperature C
62 High voltage power supply output Volts
64 Diagnostic test input (TEST_INPUT_8) mV
66 Diagnostic test input (TEST_INPUT_11) mV
68 Diagnostic temperature input (TEMP_INPUT_4) C
70 Diagnostic temperature input (TEMP_INPUT_5) C
72 Diagnostic temperature input (TEMP_INPUT_6) C
74 Ground reference (REF_GND) mV
76 4096 mV reference (REF_4096_MV) mV
78 Pre-amplified UV lamp intensity reading mV
80 1 Converter temperature C
82 5 Oxygenator flow rate cc/m
84 6 Concentration stability #2 PPB
86 7 UV lamp stability mV
100 4 TRS slope for range #1 PPB/mV
102 4 TRS slope for range #2 PPB/mV
104 4 TRS offset for range #1 mV
106 4 TRS offset for range #2 mV
108 4 TRS concentration for range #1 during zero/span
calibration, just before computing new slope and offset PPB
110 4 TRS concentration for range #2 during zero/span
calibration, just before computing new slope and offset PPB
112 4 TRS concentration for range #1 PPB
114 4 TRS concentration for range #2 PPB
116 4 TRS converter efficiency factor for range #1
118 4 TRS converter efficiency factor for range #2
MODBUS Floating Point Holding Registers
(32-bit IEEE 754 format; read/write in high-word, low-word order; read/write)
0 Maps to SO2_SPAN1 variable; target conc. for range #1 Conc. units
2 Maps to SO2_SPAN2 variable; target conc. for range #2 Conc. units
4 Maps to H2S_SPAN1 variable; target conc. for range #1 Conc. units
6 Maps to H2S_SPAN2 variable; target conc. for range #2 Conc. units
100 4 Maps to TRS_SPAN1 variable; target conc. for range #1 Conc. units
102 4 Maps to TRS_SPAN2 variable; target conc. for range #2 Conc. units
MODBUS Discrete Input Registers
(single-bit; read-only)
0 PMT detector warning
1 UV detector warning
2 Dark calibration warning
3 Box temperature warning
07266B DCN6485