Operating Instructions Teledyne API - Model T200H/T200M Operation Manual
102
In order to compensate for CPU clocks which run fast or slow, there is a variable to
speed up or slow down the clock by a fixed amount every day.
To change this variable, press:
SAMPLE ENTER SET UP PASS : 81 8
8 1 8 ENTR EXIT
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP X.X 0 ) DAS_HOLD_OFF=15.0 Minutes
NEXT JUMP EDIT PRNT EXIT
SETUP X.X SECOND ARY SETUP M ENU
COMM VARS DIAG EXIT
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL SE TUP
3x EXIT returns
to the main SAMPLE di s
p
la
y
Enter s ign a nd n umber of seconds pe r
day the clock gains (-) or loses (+).
SETUPX.X 1 ) MEASURE_MODE=NOX-NO
PREV NEXT JUMP EDIT PRNT EXIT
SETUP X. X 7) CLOCK_ADJ=0 S ec/Da
y
PREV JUMP EDIT PRNT EXIT
SETUP X. X CLOCK_A DJ: 0 Sec/Da
y
+ 0 0 ENTR EXIT
SETUP X. X 7) CLO CK_ADJ =0 Sec/D ay
PREV NEXT JUMP EDIT PRNT EXIT
Continu e to press NEXT until …