Teledyne API - Model T200H/T200M Operation Manual Operating Instructions
129
To enter the signal I/O test mode, press:
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL SETUP
EXAMPLE
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG EXIT
DIAG SIGNAL I / O
NEXT ENTR EXIT
DIAG I / O Test Signals Displayed Here
PREV NEXT JUMP PRNT EXIT
Press JUMP to go
directly to a specific
signal
See Appe ndix A-4 fo r
a complete list of
available SIGNALS
DIAG I / O JUMP TO: 5
0 5 ENTR EXIT
Enter 05 to Jump
to Signal 5 :
(CAL_LED)
DIAG I / O CAL_LED = ON
PREV N EXT JUMP ON PRNT EXIT
Exit to return
to the
DIAG menu
SETUP X.X ENTER DI AG PASS: 818
8 1 8 ENTR EXIT
Use the NEXT & PREV
keys to move between
signal types .
Pressing t he PRNT k ey will sen d a format ted print out to the seri al port and can be
captured w ith a com puter or o ther out put devic e.
EXIT
returns
to the main
SAMPLE
display
07270B DCN6512