Operating Instructions Teledyne API - Model T200H/T200M Operation Manual
130
4.13.3. ANALOG OUTPUT STEP TEST
This test can be used to check the accuracy and proper operation of the analog outputs.
The test forces all four analog output channels to produce signals ranging from 0% to
100% of the full scale range in 20% increments. This test is useful to verify the
operation of the data logging/recording devices attached to the analyzer.
To begin the Analog Output Step Test press:
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL SETUP
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXI T
SETUP X.X SECONDARY SETUP ME NU
COMM VARS DIAG EXIT
DIAG SIGNAL I / O
NEXT ENTR EXIT
Exit-Exit
returns to the
DIAG menu
SETUP X.X ENTER DIAG P ASS: 818
8 1 8 ENTR EXIT
DIAG ANALOG OU TPUT
PREV NEXT ENTR EXIT
DIAG AOUT ANAL OG OUTP UT
0% EXIT
Performs
analog output
step test.
0% - 100%
DIAG AO UT AN ALOG OUTP UT
[0%] EXIT
Pressing the key under “0%” while performing the test will
pause the test at that level. Brackets will appear around
the valu e: exam ple: [20 %] Press ing the same key aga in
will resume the test.