Operating Instructions Teledyne API - Model T200H/T200M Operation Manual
152
To manually adjust the signal levels of an analog output channel, press:
From the
AIO CONFIGURATION SUBMENU
(See Section 6.13.4.1)
DIAG ANALOG I/O CONFIGURATION
PREV NEXT ENTR EXIT
Continue pressing SET> until you reach the
output to be configured
DIAG AIO DATA_OUT_2 5V, NXCNC1, NOCAL
<SET SET> EDIT EXIT
Continue pressing SET> until ...
DIAG AIO DATA_OUT_2 CALIBRATED:NO
<SET SET> CAL EXIT
DIAG AIO DATA_OUT_2 RANGE: 5V
SET> EDIT EXIT
DIAG AIO DATA_OUT_2 VOLT-Z: 0 mV
U100 UP10 UP DOWN DN10 D100 ENTR EXIT
DIAG AIO DATA_OUT_2 VOLT-S: 4500 mV
U100 UP10 UP DOWN DN10 D100 ENTR EXIT
DIAG AIO DATA_OUT_2 CALIBRATED: YES
<SET SET> CAL EXIT
These menus
only appear if
AUTO-CAL is
turned OFF
These buttons increase /
decrease the analog output
signal level (not the value on the
display)
by 100, 10 or 1 counts.
Continue adjustments until the
voltage measured at the output
of the analyzer and/or the input
of the recording device matches
the value in the upper right hand
corner of the display (within the
tolerances
listed in Table 6-24).
DIAG AIO AOUTS CALIBRATED: NO
SET> CAL EXIT
07270B DCN6512