Teledyne API - Model T200H/T200M Operation Manual Operating Instructions
165
4.13.7.5. Flow Calibration
The flow calibration allows the user to adjust the values of the sample flow rates as they
are displayed on the front panel and reported through COM ports to match the actual
flow rate measured at the sample inlet. This does not change the hardware measurement
of the flow sensors, only the software-calculated values.
To carry out this adjustment, connect an external, sufficiently accurate flow meter to the
sample inlet. Once the flow meter is attached and is measuring actual gas flow, press:
Adjust th ese v alues
until th e display ed
flow rate equals the
flow rate being
measure d by the
independ ent fl ow
meter.
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL SETUP
SETUP X.X PRIMARY SETUP ME NU
CFG ACAL DAS RNGE PASS CLK MORE EXIT
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG EX IT
Exit at
any time
to return
to main
the
SETUP
menu
DIAG SIGNAL I /
O
NEXT ENTR EXIT
Exit returns
to the
previous menu
Repeat Pressing NE
X
T until . . .
DIAG FLOW CALIBRATION
PREV NEXT ENTR EXIT
DIAG FCAL
A
CTUAL FLOW: 480 CC / M
0 4 8 0 ENTR EX IT
ENTR ac cepts the
new value and
returns to the
previous menu
EXIT ignores the
new value and
returns to the
previous menu
DIAG FLOW SENSOR TO CAL: SAMPLE
SAMPLE OZONE ENTR EXIT
Choose between
sample and ozone
flow sensors .
SETUP X.X ENTER D IAG PASS: 8 18
8 1 8 ENTR EXIT
07270B DCN6512