Teledyne API - Model T200H/T200M Operation Manual Operating Instructions
91
To define the REPORT PERIOD, follow the instruction shown in Section 4.7.2.2 then
press:
Edit Data Channel Menu
SETUP X.X 0) CONC: ATIMER, 8, 8500
PREV NEXT INS DEL EDIT PRNT EXIT
Exits t o the mai n
Data Acquisition
menu.
SETUP X.X NAME: CONC
<SET SET> EDIT PRINT EXIT
Press SET> until you reach REPORT PERIO D (OR SAMPLE PERI OD) …
SETUP X.X REPORT PERIODD:DAYS:
0
0 0 0 ENTR EXIT
ENT
R
accepts the n ew string a nd
returns to the previous menu.
EXIT ignores the ne w strin g and
returns to the previous menu.
Press buttons to set hours
between reports in the format :
HH:MM (max: 23:59). This is a
24 hour clock . PM hours are 13
thru 23, midnight is 00:00.
Example 2:15 PM = 14:15
Set the number of days
between reports (0-366).
IIf at any time an illegal entry is selected (e.g., days > 366)
the ENTR button will disappear from the display.
SETUP X.X REPORT PERIODD:TIME:01:01
0 1 0 0 ENTR EXIT
Use the PREV and NEXT
buttons to scroll to the
data chan nel to be ed ited.
SETUP X.X REPORT PERIOD:000:01:00
<SET SET> EDIT P RIN T EXIT
From the DATA ACQUISITION menu
(see Section 6.7.2.2)
SETUP X.X ENTER DAS PASS: 818
9 2 9 ENTR EXIT
Changing t he SAMPLE
PERIOD or REPORT
PERIOD Requires a
special password