Chapter 5: Reference

The Eye Scan Tab

normal operating mode by sampling signals from the device under test.

See Also

“To set the eye scan range and resolution” on page 120

Advanced Subtab

Eye scan measurements look at selected logic analyzer channels for signals passing through small windows of time and voltage.

Advanced options let you specify the number of clocks to process in each window and whether measurement data should be accumulated or replaced.

Quick Scan Provides a faster eye scan than Complete Scan. Scanning is limited to regions that exhibit significant signal activity. It is possible that noise or other signal anomalies may not be displayed because areas that don't exhibit activity are not repeatedly scanned.

Complete Scan This is a raster scan of all pixels in the Eye Scan display. This scanning method is slower, but it ensures that noise